PREV WG 3 NEXT WG 5
IAG Earth Tide Commission (ETC)

Working Group 4:
Calibration of Gravimeters

Chairperson: M.van Ruymbeke


Report of the Working Group 4

- Status August 28, 2000 -
by M.van Ruymbeke and B.Ducarme

|||||Introduction:

|||||Description of the problem:

What kind of processes could modify the acceleration felt by the gravimeter mass? The calibration processes <2>, <3> and <4> require to move the gravimeters. Additional mechanical systems are required for processes <1> and <4>.

The calibration process <2> is generally used to calibrate an intermediate standard of the field gravimeters the so called " micrometric screw ". The instruments using a feedback system can use calibration lines to calibrate directly the feedback force if the range of the first ones does not exceed the range of the second one.

It should be pointed out that the intercomparisons in one station of several gravimeters is not a " calibration " "senso strictu".

|||||The tidal gravimeters:

Nowadays the principal types of instruments able to record tidal gravity signals are :

No calibration is required for absolute instruments(ABS) which are directly referenced to the wavelength of a laser beam and the time scale of an atomic clock. This type of gravimeter is the best one to measure the gravity values along a calibration line. Its use to record tides during a long time [Francis O., 1997] will remain marginal. It is generally used for intercomparisons with tidal instruments during a few days to calibrate them. For spring gravimeters ( LCR ) & ( ASK ), a mechanical system modulates the elastic restoring force proportionally to the rotation of a micrometer. After determination of its scale factor by intercomparisons on "calibration lines", the micrometer is used to determine the sensitivity of the gravimeters during tidal registration.
To record tides with (LCR) or other astaticized gravimeters it is necessary to use a restoring force working in the feedback mode in order to minimize the elastic after-effects inherent to the astatisation. Some (ASK) are also equipped with a system allowing to put additional masses (balls) on the beam. The equivalent force was scaled against gravity by the maker. The precision of the method is poor as it is necessary to tilt the instrument in order to put and remove the ball.

For the ( SCI ), it is not possible to modulate the feedback force directly. However the scale factor seems to be very stable [Ducarme et al., 1997]. So the " maker calibration " checked on " calibration lines " can be used for tidal records.

For ( SCG ), it is not possible to move the instruments on " calibration lines " and no internal modulation of the restoring force is possible with the required accuracy. The so-called " electrostatic calibration " gives only apparent changes of sensitivity. Direct calibration is possible only through methods <1> and <2>.

|||||Sources of errors:

The scale factor of an instrument has to be related to absolute units. During the transfer process, two kinds of errors could exist which are systematic or random. The first kind defined as systematic, is directly affecting the scale factor. This error is constant independently of the number of calibrations. The second kind which is defined like a random noise, is limiting the precision of the calibration. This kind of error decreases with the increase of the number of determinations. We can have a very high repeatability of the results of calibration, meaning a low level of random noise, associated with very large systematic errors. This risk is especially important for frequency dependent processes when the excitation periods are short like in process <4>. The slope of adjustment could be modified by an attenuation of amplitude due to low pass filtering of the mechanical or the parts of the system. As this effect is frequency dependent like the acceleration itself, the systematic errors have to be corrected by determining the transfer function of the filters. It becomes thus possible to compensate the damping of the filters at different frequencies. Systematic errors could exist in the process of transfer from the micrometer to the gravimeter itself of the scale factor ( dead zone in the mechanical transmission, long term drift, ... ). For process <1>, the gravimeters need sufficiently heavy mass to obtain significant signals. The risk exist of systematic errors induced by the mechanical effects due to the displacement of large masses. Finally it is very important to know how the calibration process itself can modify the gravimeter records, altering its sensitivity and/or drift.

|||||Selection of the methods:

It is clear that some methods are obsolete or will never reach the required accuracy. We shall try to summarize here some of the most promising approaches.
The problem is to extrapolate results obtained in the several hundred milligal range to the tidal range. It is why special base lines of a few milligal range have been established in Germany and China[ Wenzel H.G.,1995]. The second problem is to calibrate accurately the tidal records using the micrometric screw. Recent tests show that the apparent changes of sensitivity are accurately followed by the calibrations [van Ruymbeke M.,1998]. -Inertial platforms have been successfully tested [van Ruymbeke M.,1989], but much effort has still to be devoted to reach the required accuracy.

|||||Conclusions:

The 0.1% of accuracy on a phenomenon which has a so small amplitude as the tides is at the limit of instrumentation and any method to improve calibration is useful to improve the gravimeters themselves. We suggest to organize a meeting of people concerned by the determination of the scale factor of the gravimeters to overview the different approaches, including realistic evaluation of the accuracies. An intercomparison of results obtained by various ways is essential to eliminate the risk of systematic errors which are different in the various methods.

|||||Bibliography:


||||| For further information contact:

Dr. Michel van Ruymbeke
Royal Observatory of Belgium,
Av. Circulaire, 3, 1180 Brussels, Belgium
e-mail: michel.vanruymbeke@oma.be


BACK to WG's (ETC Home) PREV
WG 3
 
/
 
NEXT
WG 5